152
―――――――――――――――――――――――――――――――――――――――
11.9 Timer Trigger
―――――――――――――――――――――――――――――――――――
NOTE
・
Because triggers are not accepted while capturing data,
care must be taken over the time interval setting.
・
The start time and the stop time must be after the
START
key is pressed.
・
When the trigger mode is SINGLE, regardless of the
setting for the time interval, triggering occurs only once
at the start time.
・
In the OR setting between triggers, if the trigger other
than the timer trigger is being set, the trigger occurs
before the start time. Further, even though the stop time
is past, measurement will not be completed, because of
waiting for another trigger.
・
When trigger sources are AND, the trigger is applied
when all triggers other than the timer trigger are active
simultaneously, during the period from the start time to
the stop time. At this point, the interval setting is
disabled.