11.3 Setting the Trigger Types
―――――――――――――――――――――――――――――――――――
Trigger Screen
Waveform Screen
×
The signal ignored
0
Low level signal
1
High level signal
Selection
Trigger Screen
Method (on the trigger screen and waveform screen)
This example illustrates settings for channel A.
The channel 2 B is set in the same manner.
Using the
CURSOR
keys, move the flashing cursor, and
make the settings.
1. Set the trigger type:
Select "
LOGIC
" by using the
+
and
−
keys.
2. Set the trigger pattern:
Sets the
logic pattern for when trigger should
take place. In case of channel A, the
settings are made for logic inputs
channel A 1 to 4 in order from the
left. Select by using the
+
and
−
keys.
3. Set the AND/OR logical operator for
the trigger pattern (trigger screen
only):
Select by using the
+
and
−
keys.
This setting cannot be set on the
waveform screen.