11.1 Overview
―――――――――――――――――――――――――――――――――――
NOTE
(3) The trigger can be selected to be either the logical AND or
the logical OR combination of these trigger sources
described in
to
.
AND
: Triggering occurs only if all of the trigger
conditions hold.
OR
: Triggering occurs if any one of the trigger
conditions holds.
When the all trigger source is set to OFF, the triggering
occurs simultaneously with the start.
(4) Setting the trigger mode (except for the X-Y CONT
recorder function mode)
This determines whether or not the trigger is repeatedly
accepted after the measurement operation has
terminated.
(5) Pre-trigger and trigger timing settings
The trigger timing setting refers to which of the start and
stop events are controlled by the trigger.
The pre-trigger setting determines how much of the
captured data is actually before the trigger event.