138
―――――――――――――――――――――――――――――――――――――――
11.3 Setting the Trigger Types
―――――――――――――――――――――――――――――――――――
Rms value
Trigger level
Rms value
Trigger level
Input signal
Input signal
Trigger point
Trigger point
Upward trigger direction (
)
Downward trigger direction (
)
Trigger Screen
1 2 3
11.3.5 RMS Level Trigger (RMS recorder
function)
This trigger occurs when the input signal crosses a
predetermined trigger level (rms value) in a particular
direction ("slope":
or
). The commercial power
supplies, 50/60 Hz can be measured.
Method (on the trigger screen and waveform screen)
This example illustrates settings for channel 1(A). The
channel 2 (B) is set in the same manner.
Using the
CURSOR
keys, move the flashing cursor in the
order indicated by the numbers in the figure, and set.
1. Set the trigger type:
Using the
+
and
−
keys, select
"
RMSREV
".
2. Set the trigger level:
Using the
+
and
−
keys, set the
threshold value (voltage value).
Содержание 8804
Страница 1: ...MEMORY HiCORDER INSTRUCTION MANUAL 8804 ...
Страница 2: ......
Страница 28: ...6 1 2 System Operation ...
Страница 40: ...18 2 1 Main Units Specifications ...
Страница 90: ...68 6 3 Basic Operational Procedures 2 Press the PRINT key The manual print as the figure below is recorded ...
Страница 102: ...80 6 8 Memory Recorder Function Possible Recording Time ...
Страница 128: ...106 8 3 Basic Operational Procedures ...
Страница 200: ...178 12 4 Settings on the System Screen Page 2 ...
Страница 230: ...APPENDIX 8 Appendices Single Dual 2 Recorder function ...
Страница 232: ...APPENDIX 10 Appendices ...
Страница 237: ......
Страница 238: ......