11.3 Setting the Trigger Types
―――――――――――――――――――――――――――――――――――
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
APP
IND
Trigger point
1/2 period
Upper limit
Lower limit
11.3.4 Voltage Dip Trigger (Memory Recorder
Function)
・
The 8804 is designed to measure commercial power
supplies (50/60 Hz).
・
This unit detects momentary voltage drops in commercial
power supplies.
・
When the peak of the voltage falls lower than the setting
level, the trigger will occur.
・
The trigger levels of upper and lower limit cannot be set
separately.
・
On the trigger screen, not only the trigger level but also
the rms value is displayed.