42 |
P a g e
Step 5
Select the calibration mode, either standards
(measure prepared standards), manual (keypad
data entry), or new standards (same as Standards
but forces new standards to be measured when the
method is re-run even if history is enabled and
standards have already been measured).
Step 6 (if standards selected)
Select the number of replicates using the left and
right arrows. This determines the number of
standards to be measured and averaged at each
standard concentration point. Can be OFF (1), 2 or
3.
Step 7
Press Next
to enter the Standards screen
or press Cancel
to cancel selections and
return to the Life Science folder.
Standards Screen
Step 8
Enter the concentration values by using the keypad
numbers and the up and down arrows to move
between the different standard boxes. Range 0.001
to 9999. The C button backspaces and clears the
last digit entered.
Press the down arrow key.
Repeat until all standards are complete.
Step 9
Press Next
to enter the Calibration screen.
If there are duplicate or non-monotonic (increasing)
entries the unit will beep and highlight the incorrect
entry. Or press Back
to return to the
Parameter screen
Step 10
Select whether or not to adjust the name of your
sample before each new run, using left and arrow
keys to select Yes or No. Press
.