12 |
P a g e
Results Screen
The result at the selected wavelength is displayed on
screen.
Use the left and right arrows to move the cursor and
display the value at the cursor position (+/- 15nm from
set wavelength).
Press Cancel
to return to the Applications
Folder.
Press
to display available options which are
described below.
Options
(select using key pad numbers)
1.
Return to parameters screen (Step 1 above).
2.
Print result via selected method.
3.
Toggle between Absorbance and %T mode.
4.
Print graph – greyed out if no data are available.
7.
Sample number – add a prefix to the sample
number and reset the incrementing number to the
desired value.
8.
Save method – use the left and right arrows to
select a folder to store in (Favourites/Methods 1-9),
press the down arrow and enter name.
9.
Auto-print – toggles auto-print on/off.
Exit options by pressing