25 |
P a g e
Step 15
Press Next
when all standards have been
entered. If any duplicate or non-monotonic
(increasing entries) are present the unit will beep and
highlight the incorrect entry or press back
to
return to the Parameter screen.
Step 13
Select whether or not to adjust the name of your
sample before each new run, using left and arrow
keys to select Yes or No. Press
.
Step 14
Press
to display the replicates entry box. Use
C to clear previously stored results before measuring.
You will need to use the down arrow key to clear
each line.
Step 15
Insert the reference. Press the reference key
.
Step 16
Press
to measure the standard and store the
result.
Repeat for all replicates and standards.