Calibration Screen (Replicates off)
This shows the calibration values and allows
standards to be measured.
Step 11
Insert the reference sample. Press the reference key
. This will be used for all subsequent
samples until changed.
Step 12
Insert the standard (use C to clear previously stored
results before measuring).
Press
to measure the standard and store
the result.
Repeat step 12 for all standards. A graph will display
the results and the fitted curve as the measurements
are made.
Use the up and down arrows to select a standard to
be repeated if a poor reading has been obtained.
Use C to clear the previous reading.
Step 13
When all standards are measured the OK box
appears. Press
to accept the calibration and
go to the Results screen or press Back
to
cancel selections and return to the Parameters
screen.
Calibration Screen (Replicates on)
This shows the calibration values and allows
standards to be measured.
Step 14
Press
to display the replicate entry boxes.
Insert the reference. Press the reference key
.This will be used for all subsequent
samples until changed.
Step 15
Insert the standard and press
to measure
the standard and store the result.