23 |
P a g e
Step 5
Select the calibration mode: either Standards
(measure prepared standards), Manual (keypad data
entry), or New Standards (same as Standards but
forces new standards
to be measured when the method is re-run even if
history is enabled and standards have already been
measured)
Press the down arrow.
Step 6
Select how many replicates to have for your
experiment.
Step 7
Press Next
to enter the Standards screen or
press Cancel
to cancel selections and return
to the Applications Folder.
Standards Screen
Step 8
Enter the concentration values by using the keypad
numbers and the up and down arrows to move
between the different standard boxes. Range 0.001
to 9999.
Press the down arrow key.
Repeat step 8 for all standards.
Step 9
Press Next
when all standards have been
entered. If any duplicate or non-monotonic
(increasing entries) are present the unit will beep and
highlight the incorrect entry or press back
to
return to the Parameter screen.
Step 10
Select whether or not to adjust the name of your
sample before each new run, using left and arrow
keys to select Yes or No. Press
.