52 |
P a g e
Step 6 (if standards selected)
Select the number of replicates using the left and right
arrows. This determines the number of standards to be
measured and averaged at each standard
concentration point. Can be OFF (1), 2 or 3.
Step 7
Press Next
to enter the Standards screen or
press Cancel
to cancel selections and return to
the Life Science folder.
Standards Screen
Step 8
Enter the concentration values by using the keypad
numbers and the up and down arrows to move between
the different standard boxes. Range 0.001 to 9999. C
button backspaces and clears the last digit entered.
Step 9
Press Next
to enter the Calibration screen or
press Back
to return to the Parameter screen
Step 10
Select whether or not to adjust the name of your
sample before each new run, using left and arrow keys
to select Yes or No. Press
.
Calibration Screen (replicates off)
This shows the calibration values and allows standards
to be measured.
Step 11
Insert the reference and press the reference key
. This will be used for all subsequent samples
until changed.
Step 12
Insert the standard (use C to clear previously stored
results before measuring)
Press
to measure the standard and store the
result.