Calibration Screen (replicates off)
This shows the calibration values and allows
standards to be measured.
Step 11
Insert the reference and press the reference key
. This will be used for all subsequent
samples until changed.
Step 12
Insert the standard (use C to clear previously
stored results before measuring)
Press
to measure the standard and store
the result.
Repeat step 12 for all standards. A graph will
display the results and the fitted curve as the
measurements are made.
Use the up and down arrows to select a standard to
be repeated if a poor reading has been obtained.
Use C to clear the previous reading.
Step 13
When all standards are measured the OK box
appears. Press
to accept the calibration
and go to the Results screen (see below) or press
Back
to cancel selections and return to the
Parameters screen.
Calibration Screen
(replicates on)
This shows the calibration values and allows
standards to be measured.
Step 14
Press
to display the replicate entry boxes.
Use C to clear previously stored results before
measuring.
Step 15
Insert the standard and press
to measure
the standard and store the result.
Repeat for all replicates and standards.
A graph will display the results and the fitted curve
as the measurements are input.