27 |
P a g e
Step 20
Press
to display the replicates entry box. Use
C to clear previously stored results before measuring.
You will need to use the down arrow key to clear each
line.
Step 21
Insert the reference. Press the reference key
.
Step 22
Press
to measure the standard and store the
result.
Repeat Step 22 for all replicates and standards.
A graph will display the results and the fitted curve as
the measurements are input. Use the up and down
arrows to select a standard to be repeated if a poor
reading has been obtained. Use C to clear the
previous reading.
Step 23
Press
to accept the calibration and go to the
Results screen (see below) or press Back
to
return to the Standards screen.
Results Screen
Step 24
Insert the reference and press the reference key
. This will be used for all subsequent samples
until changed.
Step 25
Insert the sample and press
.
The concentration of the sample is taken and
displayed.
Repeat step 25 for all samples.
Press
to return to the Applications Folder.
Press
to display available Options.