53 |
P a g e
Repeat step 12 for all standards. A graph will display
the results and the fitted curve as the measurements
are made.
Use the up and down arrows to select a standard to
be repeated if a poor reading has been obtained. Use
C to clear the previous reading.
Step 13
When all standards are measured the OK box
appears. Press
to accept the calibration and
go to the Results screen (see below) or press Back
to cancel selections and return to the
Parameters screen.
Calibration Screen (Replicates on)
This shows the calibration values and allows
standards to be measured.
Step 14
Press
to display the replicate entry boxes.
Use C to clear previously stored results before
measuring.
Step 15
Insert the standard and press
to measure the
standard and store the result.
Repeat for all replicates and standards.
A graph will display the results and the fitted curve as
the measurements are input.
Use the up and down arrows to select a standard to
be repeated if a poor reading has been obtained. Use
C to clear the previous reading.
Step 16
Press
to accept the calibration and go to the
Results screen or press Back
to return to the
Parameters screen.