Calibration Screen (with Replicates off)
This shows the calibration values and allows
standards to be measured.
Step 11
Insert the reference and press the reference key
. This will be used for all subsequent
samples until changed.
Step 12
Insert the standard (use C to clear previously stored
results before measuring. Use the down arrow key to
clear each line/standard).
Press
to measure the standard and store the
result.
Repeat step 12 for all standards.
A graph will display the results and the fitted curve as
the measurements are input.
Use the up and down arrows to select a standard to
be repeated if a poor reading has been obtained. Use
C to clear the previous reading.
Step 13
Press OK
to accept the calibration and go to
the Results screen or press Back
to return to
the Standards screen.
Calibration Screen (with Replicates on, 2 or 4)
This shows the calibration values and allows
standards to be measured.
Step 14
Enter the concentration values by using the keypad
numbers and the up and down arrows to move
between the different standard boxes. Range 0.001
to 9999.
Press the down arrow key.
Repeat Step 14 for all standards.