Device Functions Proline Promag 55
7 Block OUTPUTS
Hauser
89
7.2.2
Function group OPERATION
OUTPUTS
E
⇒
CURRENT OUTPUT 1
EAA
⇓
CURRENT OUTPUT 2
EAB
⇓
PULSE/FREQ. OUTP. 1
ECA
⇒
CONFIGURATION
420
⇓
⇓
OPERATION
430
PULSE/FREQ. OUTP. 2
ECB
⇒
CONFIGURATION
420
⇓
OPERATION
430
Function description
OUTPUTS
➞
PULSE/FREQUENCY OUTPUT (1 to 2)
➞
OPERATION (FREQUENCY)
ACTUAL FREQUENCY
(4301)
!
Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to view the computed value of the output frequency.
User interface:
0 to 12500 Hz
SIMULATION
FREQUENCY
(4302)
!
Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to activate simulation of the frequency output.
Options:
OFF
ON
Factory setting:
OFF
!
Note!
• The "SIMULATION FREQUENCY OUTPUT" message indicates that simulation is
active.
• The measuring device continues to measure while simulation is in progress, i.e. the
actual measuring values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...