11.1.2
Function group OPERATION
SUPERVISION
J
⇒
SYSTEM
JAA
⇒
CONFIGURATION
800
⇓
OPERATION
804
Function description
SUPERVISION
➞
SYSTEM
➞
OPERATION
ACTUAL SYSTEM
CONDITION
(8040)
Use this function to check the present system condition.
User interface:
"SYSTEM OK" or the fault / notice message with the highest priority.
PREVIOUS SYSTEM
CONDITIONS
(8041)
Use this function to view the fifteen most recent fault and notice messages since
measuring last started.
User interface:
The 15 most recent fault or notice messages.
SIMULATION
FAILSAFE MODE
(8042)
Use this function to set all inputs, outputs and totalizers to their defined failsafe modes, in
order to check whether they respond correctly. During this time, the message
"SIMULATION FAILSAFE MODE" appears on the display.
Options:
ON
OFF
Factory setting:
OFF
SIMULATION
MEASURAND
(8043)
Use this function to set all inputs, outputs and totalizers to their defined flow-response
modes, in order to check whether they respond correctly. During this time, the message
"SIMULATION MEASURAND" appears on the display.
Options:
OFF
MASS FLOW
VOLUME FLOW
CONDUCTIVITY
Factory setting:
OFF
"
Caution!
• The measuring device cannot be used for measuring while this simulation is in
progress.
• The setting is not saved if the power supply fails.
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...