7 Block OUTPUTS
Device Functions Proline Promag 55
78
Hauser
Function description
OUTPUTS
➞
PULSE/FREQUENCY OUTPUT (1 to 2)
➞
CONFIGURATION (PULSE)
ASSIGN PULSE
(4221)
!
Note!
This function is not available unless the PULSE setting was selected in the OPERATION
MODE function (4200).
Use this function to assign a measured variable to the pulse output.
Options:
OFF
VOLUME FLOW
MASS FLOW
Factory setting:
VOLUME FLOW
Advanced options with optional software package SOLID CONTENT FLOW:
TARGET MASS FLOW
TARGET VOLUME FLOW
CARRIER MASS FLOW
CARRIER VOLUME FLOW
!
Note!
If you select OFF, the only function shown in the CONFIGURATION function group is
this function, in other words ASSIGN PULSE (4221).
PULSE VALUE
(4222)
!
Note!
This function is not available unless the PULSE setting was selected in the OPERATION
MODE function (4200).
Use this function to define the flow at which a pulse is triggered. These pulses can be
totaled by an external totalizer, and the total flow quantity since measuring started can be
registered in this way.
User input:
5-digit floating-point number [unit]: 0.0000 to 99 999
Factory setting:
Depends on nominal diameter and country (see Page 153 ff.).
!
Note!
The appropriate unit is taken from the function UNIT VOLUME (0403) or
UNIT MASS (0401), (see Page 17 or Page 15).
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...