7 Block OUTPUTS
Device Functions Proline Promag 55
70
Hauser
VALUE-f LOW
(4204)
!
Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to assign a variable to the start value frequency (4202).
The value can be higher or lower than the value assigned to the VALUE-f HIGH.
Positive and negative values are permissible, depending on the measured variable in
question (e.g. volume flow). You define a measuring range by defining the VALUE-f LOW
and VALUE-f HIGH values.
!
Note!
When conductivity is selected this value VALUE-f LOW must be 0 or positive.
User input:
5-digit floating-point number
Factory setting:
0 [unit]
!
Note!
• For graphic illustration of VALUE-f LOW see function VALUE-f HIGH ().
• The appropriate unit is taken from the function UNIT VOLUME FLOW (0402) or
UNIT MASS FLOW (0400), (see Page 16 or Page 15).
VALUE-f HIGH
(4205)
!
Note!
This function is not available unless the FREQUENCY setting was selected in the
OPERATION MODE function (4200).
Use this function to assign a variable to the end value frequency (4203).
The value can be higher or lower than the value assigned to the VALUE-f LOW.
Positive and negative values are permissible, depending on the measured variable in
question (e.g. volume flow). You define a measuring range by defining the VALUE-f LOW
and VALUE-f HIGH values.
!
Note!
When conductivity is selected this value VALUE-f HIGH must be 0 or positive.
User input:
5-digit floating-point number
Factory setting:
Depends on nominal diameter and country (see Page 153 ff.).
!
Note!
Note that values with different signs cannot be entered for VALUE-f LOW
and VALUE-f HIGH, if SYMMETRY is the setting selected for the MEASURING MODE
function (4206). In this case the message "INPUT RANGE EXCEEDED" appears on the
display.
A0001279
➀
= Value-f low
➁
= Value-f high
(continued on next page)
Function description
OUTPUTS
➞
PULSE/FREQUENCY OUTPUT (1 to 2)
➞
CONFIGURATION (FREQUENCY)
Q
125
100
0
Freq.
➀
➁
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...