10 Block SPECIAL FUNCTION
Device Functions Proline Promag 55
134
Hauser
10.1.4
Function group COATING E1
SPECIAL FUNCTION
H
⇒
ADV. DIAGNOSTICS
HEA
⇒
CONFIGURATION
750
⇓
ACQUISITION
751
⇓
CONFIG. COATING
752
⇓
COATING E1
753
Function description
SPECIAL FUNCTION
➞
ADV. DIAGNOSTICS
➞
COATING E1
REFERENCE VALUE
(7530)
Use this function to view the reference value for the decay time constant at measuring
electrode 1.
User interface:
5-digit floating-point number, including unit in milliseconds
ACTUAL VALUE
(7531)
Use this function to view the actual measured decay time constant at measuring
electrode 1.
User interface:
5-digit floating-point number, including unit in milliseconds
MINIMUM VALUE
(7532)
Use this function to view the lowest measured value for the decay time constant at
measuring electrode 1, since the last reset or deletion of the stored values.
User interface:
5-digit floating-point number, including unit in milliseconds
MAXIMUM VALUE
(7533)
Use this function to view the highest measured value for the decay time constant at
measuring electrode 1, since the last reset or deletion of the stored values.
User interface:
5-digit floating-point number, including unit in milliseconds
HISTORY
(7534)
Use this function to view the last 10 measuring values for the decay time constant at
measuring electrode 1.
User interface:
5-digit floating-point number, including unit in milliseconds
ACTUAL DEVIATION
(7535)
Use this function to view the deviation between the actual (last measured) value for the
decay time constant at measuring electrode 1 and the reference values selected in the
SELECT REFERENCE CONDITION function (7502).
User interface:
5-digit floating-point number, including unit in milliseconds
WARNING LEVEL
(7536)
!
Note!
This function is not available unless the ON setting was selected in the WARNING
MODE function (7503).
In this function, the user can specify a maximum permitted deviation (limit value) from
the reference status for the decay time constant. If this limit value is overshot or
undershot, a system error message (categorized as a notice message) is output. To do this,
the measuring system compares the actual deviation (see Function ACTUAL
DEVIATION, 7535) to the value entered here.
User input:
1 to 10000 ms
Factory setting:
100 ms
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...