(7501)
This function enables the user to start an adjustment, in order to ascertain the reference
values of various diagnostic parameters valid for his process. These reference values are
authoritative as the "starting point" for later trend analyses (regarding abrasion, corrosion
or coating formation) and should be ascertained for each process or fluid in a balanced
state.
When adjustment is performed, the reference values of the following diagnostic
parameters are ascertained:
• Decay time constant of test pulses (at measuring electrodes 1 and 2)
• Electrode potentials (of measuring electrodes 1 and 2)
• Volume flow
(flow value immediately before applying the test pulses)
Options:
CANCEL
START
Factory setting:
CANCEL
SELECT REFERENCE
CONDITION
(7502)
In this function, the reference condition is selected (at the factory or by the user), which
the affected diagnostic parameters are to be compared to later.
Options:
FACTORY (reference values determined at the factory)
USER (reference values ascertained by the user
→
Function 7501)
Factory setting:
FACTORY
WARNING MODE
(7503)
In this function, you can determine whether a warning is generated if a deviation occurs
between the reference condition (see Function SELECT REFERENCE CONDITION) and
the actual measured diagnostic parameters.
When doing so, the following diagnostic parameters are compared to the reference
condition:
• Decay time constant of test pulses
→
Function group COATING E1 or E2
• Electrode potentials
→
Function group ELECTR. POTENTIAL 1 or 2
• Volume flow
→
Function group VOLUME FLOW
Options:
OFF
ON
Factory setting:
OFF
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...