7 Block OUTPUTS
Device Functions Proline Promag 55
88
Hauser
TIME CONSTANT
(4247)
!
Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function (4200).
Use this function to enter a time constant defining how the measuring signal reacts to
severely fluctuating measured variables, either very quickly (enter a low time constant) or
with damping (enter a high time constant). Damping acts on the measuring signal before
the switch status changes, and consequently before switch-on or switch-off delay is
activated. The purpose of damping, therefore, is to prevent the status output changing
state continuously in response to fluctuations in flow.
User input:
Fixed-point number 0.00 to 100.00 s
Factory setting:
0.00 s
Function description
OUTPUTS
➞
PULSE/FREQUENCY OUTPUT (1 to 2)
➞
CONFIGURATION (STATUS)