Device Functions Proline Promag 55
7 Block OUTPUTS
Hauser
57
7.1
Group CURRENT OUTPUT (1 to 2)
7.1.1
Function group CONFIGURATION
OUTPUTS
E
⇒
CURRENT OUTPUT 1
EAA
⇒
CONFIGURATION
400
⇓
CURRENT OUTPUT 2
EAB
⇒
CONFIGURATION
400
Function description
OUTPUTS
➞
CURRENT OUTPUT (1 to 2)
➞
CONFIGURATION
ASSIGN CURRENT
OUTPUT
(4000)
Use this function to assign a measured variable to the current output.
Options:
OFF
VOLUME FLOW
MASS FLOW
CONDUCTIVITY *
* only when conductivity is switched on
Factory setting:
VOLUME FLOW
Advanced options with optional software package SOLID CONTENT FLOW:
TARGET MASS FLOW
% TARGET MASS FLOW
TARGET VOLUME FLOW
% TARGET VOLUME FLOW
CARRIER MASS FLOW
% CARRIER MASS FLOW
CARRIER VOLUME FLOW
% CARRIER VOLUME FLOW
Advanced options with optional software package ADVANCED DIAGNOSTICS:
DEVIATION COATING E1 (only when coating detection is switched on
DEVIATION COATING E2 (only when coating detection is switched on
DEVIATION ELECTRODE POTENTIAL 1
DEVIATION ELECTRODE POTENTIAL 2
DEVIATION VOLUME FLOW
DEVIATION NOISE VALUE
!
Note!
• If you select OFF, the only function shown in the function group CONFIGURATION
(400) is this function, in other words, ASSIGN CURRENT OUTPUT (4000).
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...