Device Functions Proline Promag 55
9 Block BASIC FUNCTION
Hauser
117
CONDUCTIVITY
(6405)
Use this function to switch on the conductivity measurement.
Options:
OFF
LONG INTERVAL
Measurement every 500th flow measurement
(500
×
measuring period
SHORT INTERVAL Measurement every 50th flow measurement
(50
×
measuring period
Factory setting:
OFF
!
Note!
• The availability of this function depends on characteristics of the sensors
(see COND.VITY ENABLE
Page 127).
• If conductivity is switched on it is highly recommended setting the system
damping >3
"
Caution!
Because the electrodes are not available for flow measurement during the conductivity
measurement (up to 8
×
measuring period
Page 126) the flow value will be frozen.
Therefore momentary flow changes might be unrecognized.
Function description
BASIC FUNCTION
➞
PROCESS PARAMETER
➞
CONFIGURATION