Media-Mate Plus Operation Manual
25-710-800-O
Installation
CAUTION
The Instrument should not be plugged in until installed as per the instructions below.
1.
Clean the inside of the Tank as necessary to remove any packing debris and fill with
approximately 40L of DI water (see
Section 9 – Cleaning the Tank and Fluid Lines
).
2.
Install the Heater Assembly into the Tank as shown (see
Figure 5-2
). Place Tank Cover
onto Tank (see
Figure 5.5
).
CAUTION
The filled Tank is heavy and solution may spill. This installation requires two people. The
person holding and pushing the Tank onto the Cart legs should be capable of supporting
the weight of the Tank.
3.
One person aligns the Media-Mate Plus cart legs with steps on Tank from the back. Tip
the Tank just enough so that the Cart legs can be pushed under the Tank. Lift the opposite
side of the Tank and push as you slide the Cart legs under.
HINT
: Keep the front wheels of the Cart straight.
4.
Remove the Fill Hose Assembly from its holding position on the side of the Tank and
connect to the Fill Manifold (see
Figure 5-2
).
5.
Connect the Heater connector into the Interface Panel (see
Figure 5-4
).
6.
Connect the Dispense Tubing to the Dispense Head and position over the holes in the
Tank Cover (see
Figure 5-5
).
NOTE:
Configure the Dissolution Test Station Vessel Pattern (see
Section 7 – Configuring
the Dispense Head
).
Section 5
Installation
14