Media-Mate Plus Operation Manual
25-710-800-O
Lubrication
The Fill Tubes (see
Figure 6-1
) should be lubricated every six-months to maintain easy
movement and to facilitate easy adjustment of the Volume Setting.
To Lubricate:
1.
Adjust the Volume Delivery Setting to 1000 mL (see
Section 7 – “Volume Selection”
).
2.
Apply a small amount of silicone grease (p/n 91-951-004) to the exterior of the Fill Tubes
between the Pointer and the vessel cover.
3.
Loosen the upper lock screw and rotate the Pointer in order to disengage the incremental
standard.
4.
Move the Fill Tube up and down several times between the upper and lower extreme
positions.
5.
Raise the Fill Tube to the upper position and remove excess grease.
6.
Reset Volume Selection (see
Section 7 – “Volume Section”
).
Replacing the Filter Element
A clogged Filter Element will restrict media flow and impair Instrument performance. The Filter
Element should be replaced every 90-days or whenever the Fill Level is reduced by 25 mL (this
may be observed during the Fill Cycle). Also, since the Filter Element will retain a small volume
of media, it must be replaced whenever the Tank is to be filled with a media that is different
than the media that was previously used.
Replacing the element for units without a filter housing:
1.
Unscrew the Filter Core Assembly and remove it from the Tank along with the Element
and Washers (see
Figure 9-1
).
2.
Discard the Element and replace with a new Element (p/n 91-500-099).
Figure: 9-1
Section 9
Maintenance