Media-Mate Plus Operation Manual
25-710-800-O
To Configure the Dispense Bar:
1.
For Six Vessels:
Close the Ball Valve and Pinch Valve at Vessel 7 (see
Figure 7-2
):
Figure 7-2
For Seven Vessels:
Open the Ball Valve and Pinch Valve at Vessel 7 (see
Figure 7-3
):
Figure 7-3
2.
Observe the Dissolution Test Station Vessel Pattern.
3.
Arrange the quick disconnects at the Dispense Head to match the observed pattern.
NOTE:
For six dissolution vessels, the # 7 disconnect must be connected to an inactive
location.
Section 7
Operating Instructions
28