Media-Mate Plus Operation Manual
25-710-800-O
Configuring the Dispense Head
The Media-Mate Plus Dispense Head has been designed to accommodate the Hanson SR-6,
SR-8 and SR8-Plus Dissolution Test Stations and most Test Stations manufactured by other
manufacturers. The Dispense Head is connected to the Dispense Harness by quick
disconnects which have been identified by the numbers 1 through 7. The Dispense Head will
accommodate six or seven Dissolution Vessels by enabling or disabling #7 Measure Vessel
(see
Figures 7.1
and
7.2
). The configuration is achieved by connecting the quick disconnects
so that the pattern formed matches the test station vessel pattern. Common vessel patterns
are shown below:
Typical Configurations:
3+3 Configuration
Hanson: Classic 6, SR6, SR2, SR72, QC72
VanKel: V7000, VK700
Logan: D800, D600
4+2 Configuration
Hanson: Elite 8, SR8-Plus, SR8
VanKel: VK7000, VK700
Distek: 2100A, 5100
Logan: D800, D600
4+3 Configuration
Hanson: Elite 8, SR8-Plus, SR8
VanKel: VK7000, VK700
Logan: D800, D600
4+3 Configuration
Hanson: Elite 8, SR8-Plus, SR8
VanKel: VK7000, Vk700
Logan: D800, D600
Section 7
Operating Instructions
26