Media-Mate Plus Operation Manual
25-710-800-O
When delivery volume is too low increase by lowering the Pointer as follows:
1.
Loosen the Upper Lock Screw so that the Fill Tube may be moved. Use 1/4 in. Wrench
(p/n 91-902-037).
2.
Using the Micro Drill Blank Set (p/n 92-902-036) or Calibration Gauge (91-902-045), select
a size which corresponds with the required volume change [ Volume Change X .0048 ] or
refer to the conversion chart in
Figure 6-2
.
3.
Raise the Fill Tube as necessary and place the Calibration Gauge or Drill Blank between
the Pointer and the lower side of the incremental standard notch (see
Figure 6-2
).
4.
Lower the Fill Tube so that the Pointer and Calibration Gauge or Drill Blank are tight
against the lower side of the incremental standard notch.
5.
Tighten the Upper Lock Screw and remove the Calibration Gauge or Drill Blank.
6.
Use 1/4 in. Wrench to loosen the Pointer Handle Lock Nut and slide the Pointer down tight
against the bottom of the incremental standard notch and retighten the Pointer Handle
Lock Nut.
7.
Repeat Steps 1 through 6 as necessary for each vessel.
8.
Calibrate per
Section 6 – “Volume Delivery Calibration.”
Figure 6-2
Section 6
Validation and Calibration
19