Media-Mate Plus Operation Manual
25-710-800-O
When delivery volume is too high, reduce by raising the Pointer as follows:
1.
Loosen the Upper Lock Screw so that the Fill Tube may be moved. Use 1/4 in. Wrench
(p/n 91-902 037).
2.
Using the Micro Drill Blank Set (p/n 91-902-036) or Calibration Gauge (p/n 91-902-045),
select a size which corresponds with the required volume change
[ Volume Change X
.0048 ]
or refer to the conversion chart in
Figure 6-1
.
3.
Place the Calibration Gauge or Drill Blank between the Pointer and the upper side of the
incremental standard notch (see
Figure 6-1
).
4.
Move the Fill Tube up until the Pointer and Calibration Gauge or Drill Blank are tight
against the upper side of the incremental standard notch.
5.
Tighten the Upper Lock Screw and remove the Calibration Gauge or Drill Blank.
6.
Use 3/4 in. Wrench (p/n 91-902-037) to loosen the Pointer Handle Lock Nut and slide the
Pointer up to the top of the incremental standard notch and retighten the Pointer Handle
Lock Nut.
7.
Loosen the Upper Lock Screw and move the Dispense Tube down until the Pointer is tight
against the bottom side of the incremental standard notch and retighten the Upper Lock
Screw.
8.
Repeat Steps 1 through 7 as necessary for each vessel.
9.
Calibrate per
Section 6 – “Volume Delivery Calibration.”
Figure 6-1
Section 6
Validation and Calibration
18