Media-Mate Plus Operation Manual
25-710-800-O
Start Up
CAUTION
The voltage indicated on the Media-Mate Plus System must be the same as the power source.
If the incorrect voltage is indicated, do not plug in the Instrument and contact Hanson Research
immediately.
CAUTION
Do not press the Start Switch until the indicator lights are off as this may cause overfilling of
the System.
The Instrument must be plugged into a grounded power outlet.
1.
Pull the power cord out of the retractable cord reel and plug the other end of the power
cord into a grounded three-prong power outlet.
NOTE:
The power cord plug may need to be replaced for some countries’ power outlet
configurations. Ensure that the replacement plug is approved and properly rated
for voltage and amperage by following the Power Cord Color Code below:
USA
Color Code
Wire
Description
International
Color Code
Black
Live
Brown
White
Neutral
Light Blue
Green
Ground
Green with Yellow Stripe
2.
Turn on the Main Power Switch, located on the Pump Panel next to the power cord outlet.
3.
Turn on the Air Pump and Heater power switches, located on the Control Panel.
4.
Run a Test Cycle (see
Section 7 – Test Cycle
) while observing Instrument performance.
5.
During the Test Cycle, verify the correct sequence of events by observing the Sequence
LEDs on the Control Panel and by visually observing fluid flow at the measure vessels.
6.
During the Test Cycle, verify Heater operation by observing the “1” LED on the
Temperature Controller (“1” is energized when the Heater is on).
7.
Turn off the Heater and Air Pump switches.
8.
Turn off the Main Power Switch and remove the power cord from the grounded three-
prong power outlet and return into retractable cord reel.
Section 5
Installation
15