9.2.3 Re-Installing the Front Panel
Always follow the procedure below when re-installing the front panel. Improper front panel
installation may result in damage to the instrument.
1.
Hand tighten the screws in the sequence shown in
until the
head of each screw makes contact with the front panel.
2.
Tighten screws in sequence shown in
(0.565 Newton-meter).
3.
Repeat the tightening procedure in the same sequence to 10 in.-lb
(1.125 Newton-meters).
1
Base Board
3
LCD Board
5
CPU Board
7
Opening the Front Cover
2
J4 Connector
4
J6 Connector
6
Memory Batteries
7
6
5
4
3
2
1