Basic Programming Setup
.
Status
Description
Program is Running
Data Logging, 4-20 mA outputs, sampler control and alarm checking are active.
Program is Halted
Logging stops until the program is restarted. It continues with the last logged value when restarted.
4–20 mA outputs remain unchanged
Sampler control is disabled
Alarm checking is disabled
Program is Complete or Ready to Start
No data logging
4–20 mA outputs stay at last value
No sampler interface
No alarm checking
Program Complete
A logger is off or lost power for longer than 3 hours or datalogging memory is full (see
Содержание Sigma 950