9
Section 2
Specifications
Specifications are subject to change without notice.
2.1 Factory Installed Options
General
Dimensions
34.3 cm H ´ 25.4 cm W ´ 24.1 cm D (13.5 in. ´ 10.0 in. ´ 9.5 in.)
Weight
5 kg (11 lb) not including power source
Enclosure
NEMA 4X, 6 with front cover open or closed
UV resistant, stable from -40 to 80 °C (-40 to 176 °F)
Temperature
Storage: -40 to 80 °C (-40 to 176 °F) Operate: -10 to 65.5 °C (14 to 150 °F)
Power Requirements
12 V dc supplied from 6 amp-hr. gel or 4 amp-hr. Ni-Cad battery
115 V ac, 230 V ac or 100 V ac power supply/battery charger
Graphics Display
Back-lit liquid crystal display (LCD), auto-off when not in use (under battery operation).
8 line ´ 40 character in text mode, 60 ´ 240 pixels in graphics mode.
Keypad
21-position sealed-membrane switch with blinking green LED to indicate power on. Four “soft
keys,” functions defined by display.
Totalizers
8-digit resettable and 8-digit non-resettable software
Time Base Accuracy
± 0.007%
Measurement Modes
Flumes: Parshall, Palmer Bowlus, Leopold-Lagco, H, HL, HS, Trapezoidal
Weirs: V-notch, Contracted/Non-Contracted rectangular, Thel-Mar, Cipolletti
Manning Equation: Round, U, and Rectangular Channels
Flow Nozzle: California Pipe
Head vs. Flow: Custom programmable curve of up to 100 points.
Level only: Inches, feet, centimeters, meters
Area Velocity: Level-area table, Circular pipe, U-shaped channel, Trapezoidal channel,
Rectangular channel
Power Equation:
Data Logging
Memory Mode: Either slate or wrap-around may be selected.
Data Points: Approximately 20,000 standard. Expandable up to 116,000 data points.
Daily statistics: Available for up to 32 days.
Recording Intervals: 1, 2, 3, 5, 6, 10, 12, 15, 20, 30, 60 minutes
Sampler Output
12 to 17 V dc pulse, 100 mA max at 500 ms duration
Communications
RS232 Serial Interface, SCADA-Modbus
®
ASCII communication protocol
Integral pH/Temperature/ORP Meter (pH and ORP cannot be simultaneously measured)
Control/Logging
Field selectable to log pH/temperature/ORP independent of flow or in conjunction with flow;
also controls sample collection in response to value of low/high setpoints.
pH/Temperature Sensor
Temperature compensated; impact resistant ABS plastic body; combination electrode with
porous Teflon junction.
Measurement Range
2 to 12 pH, within specifications, 0–14 plt. maximum range.
Operating Temperature
-10 to 80 °C (14 to 176 °F)
Dimensions
19.5 mm D ´ 15.24 cm L (0.75 in. ´ 6 in.) with 19.5 mm (0.75 in.) mpt cable end
Totalizers
6-digit non-resettable mechanical Units: ft³, gal, m³, liter, acre-ft
Q
K
1
H
n
1
K
2
H
n
2
+
=
Содержание Sigma 950
Страница 2: ......
Страница 6: ...4...
Страница 20: ...18 Introduction...
Страница 34: ...32 Basic Programming Setup...
Страница 50: ...48 Sensor Installation...
Страница 82: ...80 Maintenance...
Страница 83: ...81 Appendix A Program Flow Charts Figure 22 Overview of Basic Program Menus...
Страница 84: ...82 Program Flow Charts Figure 23 Setup Flow Chart...
Страница 85: ...83 Program Flow Charts Figure 24 Options Flow Chart...
Страница 86: ...84 Program Flow Charts Figure 25 Alarms Menus Flow Chart...
Страница 87: ...85 Program Flow Charts Figure 26 Calibration Menus Flow Chart Page 1...
Страница 88: ...86 Program Flow Charts...
Страница 102: ...100 Programming Features...
Страница 106: ...104 Primary Devices Head Measurement Locations...
Страница 126: ...124 SCADA Modbus System Guidelines...