62
Communications Setup
1.
Press
CHANGE CHOICE
until the desired reporting method is displayed then press
ACCEPT
.
2.
When the 950 Flow Meter calls the pager service, it will transmit a
Pager Alarm Code number (see
) which corresponds to a specific alarm
condition.
11:00 AM 21 - APR - 01
MODEM SETUP
ACCEPT
REPORTING ORDER:
CHANGE
CHOICE
PAGER THAN MODEM
CANCEL
CHOICES: MODEM AND / OR PAGER
Table 24 Pager Alarm Codes
Alarm
Code #
Reason
Low Main Battery
1
Battery pack is less than 11.5 V
Memory Battery
2
Internal memory battery is low
Low Slate Memory
3
Less than 10% slate memory left
Slate Memory Full
4
Slate memory is used up
—
6
Reserved for Sampler
—
7
Reserved for Sampler
—
8
Reserved for Sampler
—
9
Reserved for Sampler
U-Sonic Echo Loss
10
No return signal detected
Xducer Ringing
11
The return signal is detected too soon
U-Sonic failure
12
Ultrasonic board detects an error
RS485 Timed Out
13
Comm. problems with RS485
—
14
Reserved for Sampler
—
15
Reserved for Sampler
Low Bubbler Pres.
16
Possible leak in bubble tank
Clogged Bubbler
17
Bubbler tube is plugged
High Level
18
—
High Flow
19
—
High Flow Rate of Chg.
20
—
High pH/ORP
21
—
High Process Temperature
22
—
High Rainfall
23
—
High CH1
24
—
High CH2
25
—
High CH3
26
—
High CH4
27
—
High CH5
28
—
High CH6
29
—
High CH7
30
—
Содержание Sigma 950
Страница 2: ......
Страница 6: ...4...
Страница 20: ...18 Introduction...
Страница 34: ...32 Basic Programming Setup...
Страница 50: ...48 Sensor Installation...
Страница 82: ...80 Maintenance...
Страница 83: ...81 Appendix A Program Flow Charts Figure 22 Overview of Basic Program Menus...
Страница 84: ...82 Program Flow Charts Figure 23 Setup Flow Chart...
Страница 85: ...83 Program Flow Charts Figure 24 Options Flow Chart...
Страница 86: ...84 Program Flow Charts Figure 25 Alarms Menus Flow Chart...
Страница 87: ...85 Program Flow Charts Figure 26 Calibration Menus Flow Chart Page 1...
Страница 88: ...86 Program Flow Charts...
Страница 102: ...100 Programming Features...
Страница 106: ...104 Primary Devices Head Measurement Locations...
Страница 126: ...124 SCADA Modbus System Guidelines...