93
Programming Features
Memory can be configured in slate or wrap mode.
Note:
When slate memory mode is used and becomes full, the 950 will enter program complete
mode and stop logging data.
Slate Memory Mode
—Slate mode causes logging to stop when memory becomes full.
The flow meter continues to operate but no more data is logged. Use this mode so no
data is lost from the beginning of the logging period.
Wrap Memory Mode
—In wrap mode, when memory becomes full, the oldest reading is
discarded each time a new reading is taken. When memory becomes full, the flow meter
continues to operate and log data. This mode is best used to receive the most recent
data readings.
Memory Mode Configuration
1.
Select
DATA LOG
from the Advanced Options menu.
2.
Select
MEMORY MODE
using the
UP
and
DOWN
soft keys. Press
ACCEPT
.
3.
Press
CHANGE CHOICE
to pick either Slate or Wrap. Press
ACCEPT
.
B.14 Datalogging Configurations
1.
From the Main Menu, select
OPTIONS>ADVANCED OPTIONS> DATA LOG
.
2.
Highlight
SELECT INPUTS
using the
UP
and
DOWN
soft keys. Press
SELECT
.
3.
Highlight the channel to log using the
UP
and
DOWN
soft keys. Certain channels
require more information.
4.
Press
CHANGE CHOICE
to select Logged or Not Logged. Press
ACCEPT
.
5.
Enter a logging interval using the numeric keypad, then press
ACCEPT
. Valid logging
intervals are shown on the status bar along the bottom edge of the display.
6.
Select another channel to configure or press
RETURN
to back up one step.
B.15 Diagnostics
From the Main Menu, select
OPTIONS > ADVANCED OPTIONS > DIAGNOSTICS.
In addition to the automatic diagnostics that are performed upon power up, a keypad test,
LCD test, demonstration graph, velocity analysis, and events (log) are available.
Table 37 Setup Parameters for Specific Channels
Channel Name
Configuration Options
Process Temperature
1.
Select Logged or Not Logged using the
CHANGE CHOICE
soft key.
2.
Press the
ACCEPT
soft key to continue.
3.
Enter the Logging Interval using the numeric keypad.
4.
Select Temperature Units, °F or °C (this is the only place where temp. units can be changed).
Rainfall
pH/ORP
and
Level / Flow
Analog Inputs
Содержание Sigma 950
Страница 2: ......
Страница 6: ...4...
Страница 20: ...18 Introduction...
Страница 34: ...32 Basic Programming Setup...
Страница 50: ...48 Sensor Installation...
Страница 82: ...80 Maintenance...
Страница 83: ...81 Appendix A Program Flow Charts Figure 22 Overview of Basic Program Menus...
Страница 84: ...82 Program Flow Charts Figure 23 Setup Flow Chart...
Страница 85: ...83 Program Flow Charts Figure 24 Options Flow Chart...
Страница 86: ...84 Program Flow Charts Figure 25 Alarms Menus Flow Chart...
Страница 87: ...85 Program Flow Charts Figure 26 Calibration Menus Flow Chart Page 1...
Страница 88: ...86 Program Flow Charts...
Страница 102: ...100 Programming Features...
Страница 106: ...104 Primary Devices Head Measurement Locations...
Страница 126: ...124 SCADA Modbus System Guidelines...