Free Chlorine
50081_Chlorine_Free_Forward.fm
Page 93
Free Chlorine
c
• To avoid loss of chlorine, be careful not to agitate the sample when measuring or pouring.
• Sample glassware must be cleaned and pretreated to satisfy any chlorine demand before the sample is introduced. See
for additional information.
• Always use organic-free water for sample dilution (Cat. No. 26415-49).
• When sampling tap water, let the faucet run for at least 4–5 minutes prior to collecting the sample.
• A convenient test sample can be prepared by diluting Chlorine Standard Solution (Cat. No. 14268-10) with deionized (DI) water.
• Rinse the electrode, temperature probe, and anti-diffusion tip with deionized water before every titration.
• Download the “Certificate of Analysis” (COA) to obtain the exact concentration of any unopened bottle of Hach titrant or standard
solution. See
Obtaining the Certificate of Analysis for Hach Reagents
.
• Hach buffer reagents for chlorine titrations are highly recommended for this analysis.
• Never substitute buffers designed for calibrating pH meters. They may contain dyes that interfere with amperometric titration.
• Never use buffers contaminated with mold or bacteria.
• Adjust the contrast of the display by accessing the Methods menu and then pressing
7
to lighten or
0
to darken.
• A distinction is drawn between running a new test and a new sample. Each test is a replicate run of the current or previous
sample analysis. The AutoCAT automatically tracks the results of a series of tests, and automatically calculates the mean and
standard deviation for all the results. Each test is given a different number, all of which are listed under the same Sample ID.
When starting a new sample, a new ID is assigned.
• The AutoCAT can accommodate any sample size.
Standard Methods
recommends a default volume of 200 mL. The volume can
be adjusted to expand the test range.
• The analysis data may be analyzed repeatedly from the Zoom Range Determination window, in manual (MEPD) and/or automatic
(AEPD) modes. The results are not stored until accepted.
• The AutoCAT calculates the chlorine concentration based on the sample volume. Make sure that the sample volume is correct.
• Press the
STOP
key any time to interrupt instrument operation. The user will be given the option to resume, restart, or quit.
• Press the
CHECK MARK
any time during data acquisition to halt data collection.
• Press the
DEL
key during any timer period (except Start Timer) to immediately exit the timer and proceed to the next step in
the procedure.
• Flush the burette each day before the first sample test or calibration is performed. See Daily Start-up on page 26.
• Flush the burette when changing titrants. See Changing Titrants on page 25.
Amperometric Forward Titration
Titration Workstation
0.100–5.000 mg/L as Cl
2
*
Scope and Application:
For drinking water; USEPA accepted for reporting**
* Higher ranges can be determined through sample dilution. See
.
**Procedure equivalent to USEPA method 330.2 and
Standard Methods
4500-Cl.D. for drinking water.
Sample Tips and Techniques
Reagent Tips and Techniques
Instrument Tips and Techniques
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......