Troubleshooting
Troubleshooting
50081_Troubleshooting.fm
Page 143
General
quality of the
Titration
Curve
The end point
is not sharp.
If the electrode hasn't
been used for some time
it may need to be
reconditioned.
• Clean and recondition the electrode. See
The titration
curve is
unusually
noisy.
Mechanical vibrations of
the sample cell or
instrument increases the
noise in the collected
titration curve.
• Remove any source of vibration. This includes repositioning
the sample beaker to avoid having the magnetic stir bar strike
it's side.
• Ensure that the plastic anti-diffusion tip and electrode are
properly installed into the electrode head. See
.
Sharp spikes
are observed
in the titration
curve.
This is generally not a
major problem but
electrical spikes may be
observed when the
AutoCAT shares an
electrical circuit with other
devices or relays during
the data collection
process. Even if there are
electrical spikes in the
titration curve, they
generally don't prevent
the analysis of the data.
• Install a surge protector between the AutoCAT 9000 and the
wall outlet.
• Turn OFF all other devices on the same circuit.
• Use only a dedicated electrical circuit for the AutoCAT 9000.
Erratic
operation
The AutoCAT 9000 is
sensitive to static
electricity.
• Surround the AutoCAT 9000 with a Faraday cage.
• Move the instrument to a location free of strong electrical
fields or static discharges if performance is
adversely affected.
Irregular
distortions or
flat regions in
the middle of
the titration
curves.
Occasionally air bubbles
get trapped in the system
and pass through burette
tip. If air bubbles pass
through the burette tip
before the titration has
reached the end point
then the apparent titrant
volume will be larger than
it really should be. Air
bubbles may also develop
when the titrant degasses
due to temperature
changes and/or pressure
drops within the system.
• Inspect the tubing and remove or purge any air bubbles that
might be trapped in the system; also check to be sure all the
tubing connections are tight and not leaking. Loose
connections are the primary source for recurring air
bubbles to enter the system.
The titration
curve starts or
ends with a
high, perfectly
flat region.
The current range for the
A/D converter has been
exceeded.
• Increase the current range.
• Predose the sample with enough titrant to bring the signal
into the appropriate current range.
• Use a proportionately smaller sample size to bring the signal
into the appropriate current range.
Problem
Category
Symptom
Probable Cause
Corrective Action
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......