Sulfite
Sulfite
50081_Sulfite_Back.fm
Page 113
9.
Highlight Oxidant vol.
Press the
CHECK MARK
to
enter a new oxidant volume
(see Table 1), or proceed to
.
Note:
If the sample has not
been fixed this procedure
requires the addition of oxidant
prior to analysis.
10.
Highlight Volume for
Test 1.
Press
1
to confirm.
Press the
CHECK MARK
to
enter a new sample volume.
11.
Add 1 SwifTest
dispensation (0.1 g) of
potassium iodide (KI).
Press the
CHECK MARK
to
confirm.
Note:
KI is added in excess.
The precise amount added is
not crucial for the accuracy or
precision of the analysis.
12.
Add one dose of pH 7.0
Phosphate Buffer.
Press the
CHECK MARK
to
confirm.
Note:
Phosphate buffer is
added to adjust the sample pH.
The amount used is not crucial
for the accuracy or precision of
the analysis.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......