Chlorite
Chlorite
Page 130
50081_Chlorite.fm
17.
Select type of end point
determination.
Press
DEL
for Manual.
Note:
Automatic determination
attempts to identify linear
segments without user input is
not recommended
for this
method.
18.
Use the navigation keys
to select two line segments,
one on each side of the
titration end point, by
positioning vertical lines
(cursors) at the beginning
and end of each segment.
Press the
CHECK MARK
to
confirm changes or press
DEL
to return to the Zoom
window.
Note:
For more information on
manual EP determination, see
page 135.
19.
Press the
CHECK MARK
to accept the analysis and
proceed to Titration 2, or
press
DEL
to return to the
previous window to adjust
the linear segments and
reanalyze the data.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......