Total Oxidants
Total Oxidants
50081_TotalOxidants.fm
Page 121
9.
Add 1.0 g of potassium
iodide (KI).
Press the
CHECK MARK
to confirm.
Note:
KI is added in excess;
the precise amount added is
not crucial to the accuracy or
precision of the analysis.
10.
Add 1.0 mL of pH 4
acetate buffer.
Press the
CHECK MARK
to confirm.
Note:
Acetate buffer is added
to adjust the sample pH; the
precise amount added is not
crucial to the accuracy or
precision of the analysis.
11.
The timer will start.
This allows a short period of
time for the reagents to mix
before starting the titration.
12.
Lower the electrode
head onto the beaker rim.
Press the
CHECK MARK
to confirm.
13.
The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
CHECK MARK
to change one
of the analysis parameters or
press 1 to confirm.
Note:
Any changes made will
become the new default
parameters for this method
Sub-ID.
Note:
These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note:
This step will be skipped
if autoscaling software is
enabled.
14.
The timer will start,
allowing the electrode to
stabilize before data
collection.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
15.
The Zoom Range
Determination window will
display the titration curve.
Use the navigation buttons to
select and move the cursors
to define the range of data
points for analysis.
Note:
Defining the range of
data points excludes
undesirable portions of the
titration curve from
the analysis.
16.
Select type of end point
determination.
Press
DEL
for Manual.
Note:
Automatic determination
attempts to identify linear
segments without user input
and is
not
recommended
for
this method.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......