![Hach AutoCAT 9000 Скачать руководство пользователя страница 139](http://html1.mh-extra.com/html/hach/autocat-9000/autocat-9000_instruction-manual_3154724139.webp)
Below Detectable Limit
50081_BDL.fm
Page 137
Below Detectable Limit
Method Limits
When a titration result is below the limits of the Hach method in use, the instrument
will return a result of “Below Detectable Limit” (BDL). During normal operation, drift
can occur. Even when a result reads below the method threshold, a real endpoint with
low noise level is possible.
The BDL feature allows for review of the titration curve for a usable endpoint.
Example 1 shows a BDL result with a high level of noise and no clear endpoint.
Example 2 shows a BDL result with usable data (low noise and a sharp endpoint).
Example 1
Example 2
1.
Press
DEL
to view the
titration curve.
2.
Press the
CHECK MARK
to accept BDL as the result.
3.
In this example, “Below
Detectable Limit” is stored
with Test 1 for Sample 1.
1.
Press
DEL
to view the
titration curve.
Example 2, continued
2.
Press
DEL
to perform
manual endpoint
determination.
3.
Use the navigation keys
to select two line segments,
one on each side of the
titration end point. Position
the vertical lines (cursors) at
the beginning and end of
each segment.
Press the
CHECK MARK
to
confirm changes or press
DEL
to return to the Zoom
window.
4.
Press the
CHECK MARK
to accept this calculated
result, press
DEL
to return to
the previous window to
adjust the linear segments
and analyze the data again.
5.
If the results are not
usable, press
DEL
twice to
return to the initial Zoom
Range Determination
screen. Press the
CHECK MARK
to accept
BDL.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......