Maintenance
50081_Maintenance.fm
Page 139
Section 9
Maintenance
9.1 Instrument Maintenance
9.1.1 General
Protect the instrument from temperature extremes including direct sunlight, heaters, and
other heat sources. Do not store or use the instrument in a dusty, damp, or wet location.
9.1.2 Display
Do not scratch the display screen. Never use a pen, pencil, or other sharp object on the
display surface. The display is water-resistant, but should not be exposed to direct rain or
moisture. The display screen contains a glass element. Protect the screen against
strong impacts.
9.1.3 Cleaning the Instrument
Use a soft, damp cotton cloth to wipe the instrument and the display screen. If the surface
of the screen becomes soiled, cleaned it with a soft cloth and diluted window cleaning
solution. Do not use strong solvents.
9.1.4 Cleaning and Conditioning the Electrode
Clean and condition the AutoCAT 9000™ dual platinum electrodes (DPE) as shown in
the procedure below, to make them more responsive. The frequency of use and the type
of samples analyzed will dictate the frequency for electrode cleaning and conditioning.
Excess “noise” in the titration curve and/or poor curve shapes can indicate that the
electrode is dirty.
1.
Remove the plastic anti-
diffusion dispenser tip from
the electrode head and set
aside. See
for
more information.
2.
Prepare a dilute nitric
acid solution. Measure
approximately 200 mL of
solution into a clean
250-mL beaker.
Note:
Hach offers 1:1 (v/v)
Nitric Acid (Cat. No. 2540-49)
for electrode cleaning.
3.
Turn the instrument ON
or, if already in operation,
access the
Electrodes
menu.
Note:
The electrode cleaning
routine is identical for all
methods.
4.
Press
5
to select
Electrode Cleaning, and
follow the embedded
procedure.
After electrode cleaning is
complete, replace the plastic
anti-diffusion tip to the
electrode head.
Note:
Electrodes should be
conditioned prior to first
running a titration. To condition
the electrode, soak in a dilute
chlorine solution.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......