Sulfite
Sulfite
50081_Sulfite_Back.fm
Page 117
Summary of Method
This procedure follows method number 4500-SO
3
2–
B in
Standard Methods for the
Examination of Water and Wastewater
. In the amperometric back titration, a measured
volume of oxidant (iodine) is added directly to a measured volume of the sample at
collection time. This “fixes” the sulfite concentration and decreases the amount of
sample loss throughout the titration. The method is performed at neutral pH to minimize
interference from dissolved oxygen and organic chloramines.
At the time of analysis, the sample is adjusted to pH 7 with phosphate buffer, and excess
potassium iodide is added. The unreacted oxidant is then titrated with standard PAO
solution. The sulfite concentration is derived from the difference in the amount of
oxidant originally added to the sample and the amount remaining prior to titration.
The chemical reactions are:
Accounting for Sample Iodine Demand
Iodine demand in the sample can occasionally have a significant impact on the accuracy
of a titration. This can be accounted for by titrating the sample matrix prior to sulfite (or
sulfur dioxide) addition. For best results, chlorine should be eliminated from the sample
by UV irradiation. The resulting value from the titration is the blank value, and should be
performed at least once daily when sulfite titrations are run. The blank value is
subtracted from the result of each sulfite titration for maximum accuracy.
Required Reagents
Description
Qty. required per test
Unit
Cat. No.
Iodine Standard Solution, 0.0282 N ............................................................ varies................ 1000 mL .............23333-53
Phenylarsine Oxide (PAO) Titrant, 0.00564 N............................................ varies................ 1 L ........................1999-53
Phosphate Buffer, pH 7, w/dropper ............................................................. 1 mL................. 100 mL ...............21553-32
Potassium Iodide, SwifTest refill ................................................................0.10 g ............... 25 g.......................1077-60
Required Apparatus
Beaker, Glass, 250-mL ................................................................................ 1 ....................... each......................500-46H
Cylinder, Graduated, 250 mL ......................................................................1 ....................... each.........................508-46
Magnetic Stir Bar, Teflon
®*
coated............................................................. 1 ....................... each.....................50085-00
SwifTest dispenser.......................................................................................1 ....................... each.....................28341-00
Optional Reagents
Chlorine Standard Solution, Voluette
®
Ampules ........................................ varies................ 16/pkg.................14268-10
Dilution Water, organic-free........................................................................ varies................ 500 mL ...............26415-49
Phenylarsine Oxide (PAO) Titrant, 0.00564 N, cartridge ........................... varies................ each.......................1949-01
Sodium Thiosulfate Standard Solution, stabilized, 0.00564 N ...................varies................ 1 L ......................24088-53
I
3
–
SO
3
2–
H
2
O
+
+
3I
–
HSO
4
–
H
+
+
+
→
PhAsO PAO
(
)
I
3
–
2H
2
O
+
+
3I
–
PhAsO OH
(
)
2
2H
+
(Ph=Phenyl)
+
+
→
*Teflon is a Registered Trademark of EI DuPont Company.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......