Iodine
Iodine
Page 60
50081_IodineCalibration.fm
17.
The calibration titration
starts by filling the burette
with titrant and
subsequently dispensing a
3.4-mL predose of Iodine
into the calibration sample.
This will bring the electrical
signal into measurable
range.
18.
The Start Timer will
begin to allow the electrode
to stabilize before data is
collected.
After the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
19.
The Zoom Range
Determination window will
display the titration curve.
Select and move the vertical
lines (cursor) to define the
range of data points for
analysis.
Note:
Defining the range of
data points eliminates
undesirable portions of the
titration curve.
Select type of end point
determination:
Press the
CHECK MARK
for
Automatic and proceed to
or press
DEL
for
Manual and proceed to
.
20.
Select two line
segments, one on each side
of the titration end point, by
positioning the cursors at
the beginning and end of
each segment.
Press the
CHECK MARK
to
confirm changes.
Press
DEL
to return to the
previous window.
21.
Press the
CHECK MARK
to accept the analysis.
Proceed to
.
Press
DEL
to return to the
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
22.
The AutoCAT will run
the analysis based on the
automatic selection of linear
segments.
Note:
This algorithm requires
approximately 60 seconds to
complete the analysis.
23.
Press the
CHECK MARK
to accept the analysis.
Press
DEL
to return to the
previous window (Zoom
range determination) to
adjust the linear segments
and reanalyze the data.
24.
Press
1
to select
Continue.
Press
2
to select More Details
and view analysis statistics
specific to the current
calibration.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......