Collecting and Accessing Data
Collecting and Accessing Data
Page 46
50081_AutomatedTitration.fm
5.2.3.3 Dilution
If a sample solution is diluted prior to analysis, the sample parameter Dilution should be
enabled (set to Yes), allowing access to three other parameters (Predilution vol., Final
dilution vol., and Sample vol.). These parameters allow the calculated analyte
concentration to be corrected for dilution and are only applicable when dilution is
enabled.
This is generally done to bring the analyte concentration into the proper range for
analysis. The AutoCAT 9000 can automatically correct any measured analyte
concentration for sample dilution and three parameters are used to calculate analyte
concentration:
•
Predilution volume is the volume of the original sample
•
Final dilution vol. is the volume of the sample after dilution
•
Volume for Test is the volume of solution taken from the diluted sample
and analyzed.
Normally, AutoCAT returns the concentration of the sample actually presented to it (i.e.,
the concentration of the sample placed in the analysis beaker and analyzed by the
instrument). Using the dilution information, the AutoCAT can automatically convert the
concentration measured for the diluted sample into the concentration of the undiluted
sample.
5.2.3.4 Reductant Concentration
When performing a back titration the sample is fixed by spiking it with reductant (either
PAO or sodium thiosulfate) solution. Reductant concentration is the concentration (mN)
of the solution used to fix the sample. The exact concentration of a Hach reductant
solution can be obtained. See
Obtaining the Certificate of Analysis for Hach Reagents
5.2.3.5 Reductant Volume
In a back titration the sample is fixed by spiking it with reductant (either PAO or sodium
thiosulfate) solution. Reductant Volume is the volume (mL) of the reductant solution
used to fix the sample.
5.2.4 Analysis Calculations
When two or more consecutive new tests (as opposed to new samples) are run, the
AutoCAT 9000 automatically calculates the mean and the standard deviation of the mean
for the results obtained from the individual tests. This is equivalent to making several
replicate measurements starting with identical samples for each measurement. The
standard deviation of the mean is different than the standard deviation of the replicate
measurements.
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......