![Hach AutoCAT 9000 Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/hach/autocat-9000/autocat-9000_instruction-manual_3154724105.webp)
Total Chlorine
Total Chlorine
50081_Chlorine_Total_Back.fm
Page 103
Note:
These recommendations apply to Total Chlorine back titrations, correspond to data
collection times typically under 5 minutes, and provide adequate data points on both sides of
the titration end point for proper analysis. Adjusting the volume increment will reduce the
data collection times and number of data points collected. For example, doubling the volume
increment will cut the data collection time and the number of points collected in half.
Method Performance
Estimated Detection Limit
Based on the USEPA procedure of estimating the method detection limit
(40 CFR Pt.136, App. B, 7-1-95), the estimated detection limit, using the procedure as
written, is 5.1 µg/L (0.0051 mg/L) chlorine.
Safety
Use good safety practices and laboratory techniques throughout the procedure. Consult
the Material Safety Data Sheet (MSDS) for specific reagent(s) information.
Table 1 Recommended Volume Increment Settings for Total Chlorine (Back)
Estimated Chlorine Conc. (mg/L)
Suggested Volume Increment (mL)
<0.10
0.0010*
0.10–0.90
0.0030
0.90–1.00
0.0010
* with pre-dose
Table 2 Precision Statements
At Concentration
95% Confidence Limits
0.023 mg/L Cl
2
0.020–0.026 mg/L Cl
2
0.398 mg/L Cl
2
0.370–0.426 mg/L Cl
2
1.313 mg/L Cl
2
1.287–1.339 mg/L Cl
2
4.751 mg/L Cl
2
4.689–4.813 mg/L Cl
2
Table 3 Accuracy Assessments
Sample
Spike Addition
Average % Recovery
% Range of
Recovery
Drinking Water
Free Cl
2
at 0.465 mg/L
97.3 (
n
=4)
95.8–99.1
Denitrified
Wastewater
Effluent
(Source #1)
NH
2
Cl at 0.423 mg/L
87.23 (
n
=3)
73.5–94.8
NH
2
Cl at 0.844 mg/L
93.07 (
n
=3)
91.8–93.8
NH
2
Cl at 1.262 mg/L
88.56 (
n
=3)
85.3–91.9
Denitrified
Wastewater
Effluent
(Source #2)
NH
2
Cl at 0.423 mg/L
92.59 (
n
=3)
88.9–97.2
NH
2
Cl at 0.844 mg/L
98.06 (
n
=3)
96.8–98.8
NH
2
Cl at 1.262 mg/L
95.01 (
n
=3)
89.9–98.0
Содержание AutoCAT 9000
Страница 2: ...50081_TTL fm...
Страница 8: ...Page 6 50081_Safety fm...
Страница 14: ...Page 12 50081_Introduction fm...
Страница 24: ...Page 22 50081_OperationStop fm...
Страница 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Страница 44: ...Page 42 50081_SoftwareSetup fm...
Страница 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Страница 52: ...Page 50 50081_CalibrationStopper fm...
Страница 66: ...Page 64 50081_Sampling fm...
Страница 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Страница 68: ...Page 66 50081_ProcedureStopper fm...
Страница 138: ...Page 136 50081_ManualEndpoint fm...
Страница 140: ...Page 138 50081_BDL fm...
Страница 154: ...Page 152 50081_Titration fm...
Страница 158: ...Page 156 50081_PrinterInstallation fm...
Страница 160: ...Page 158 50081_ReplacementParts fm...
Страница 170: ...Page 168 50081_GeneralInfoStop fm...
Страница 176: ......