safety, warnings, precautions, prohibitions
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 20 of 213
Pull-cord
actuator (lanyard
with
magnetic switch
on the
UserTerminal and
waist belt on the
subject)
N
magnet is pulled off through
the cord
N
running belt & inclination
system stop fast
N
machine is NOT cut off from
main power of drive motor
and inclination system
N
display active and shows
“PULL STOP”
N
interface active, but no start,
speed or elevation signal
executed
N
keyboard inactive
communication
NOT interrupted
during
“pull stop” stage
N
reconnect magnet
to its position on
UserTerminal
N
press START key,
N
or new command
through interface
Additional stop
button for arm
support,
incorporated on
top of the
handgrip
N
Stop button is pushed
N
running belt & inclination
system stop fast
N
machine is NOT cut off from
main power of drive motor
and inclination system
N
display active and shows
“PULL STOP”
N
interface active, but no start,
speed or elevation signal
executed
N
keyboard inactive
communication
NOT interrupted
during
“pull stop” stage
N
release push force
from the button
N
press START key,
N
or new command
through interface
quick stop
through light
barriers
at belt re-entry
zones, only at
oversize model
families
200-450/75-300
(venus/saturn).
N
light beam is interrupted
N
running belt & inclination
system stop extremely fast
N
drive motor is used to stop
the running belt very fast
N
machine is NOT cut off from
main power of drive motor
and inclination system
N
display active and shows
“PULL STOP”
N
interface active, but no start,
speed or elevation signal
executed
N
keyboard inactive
communication
NOT interrupted
during Quick-
Stop stage
N
remove obstacle
from light beam
N
press START key,
N
or new command
through interface
In all above mentioned cases all displays will reset to the value “0”, independent from the type of stop function
and independent from the way, if the machine was cut off from main power or if the main power remained.