[12.M2]
Change from forward to reverse rotation
N
Turn the key-operated switch and start the machine without a subject on the running belt only when the belt is
not moving.
N
Observe the belt position for 2 minutes and correct the belt position with the belt adjustment screw (see
chapter maintenance and belt adjustment). It is normal, that after changing the belt rotation, the belt will move
to one side and therefore the running belt needs to be readjusted to the center of the running track.
N
Once the belt is readjusted, STOP the belt and operate the machine with the subject on the belt.
[12.M3]
Change from reverse to forward rotation
N
Turn the key-operated switch and start the machine without a subject on the running belt only when the belt is
not moving.
N
Observe the belt position for 2 minutes and correct the belt position with the belt adjustment screw (see
chapter maintenance and belt adjustment). It is normal, that after changing the belt rotation, the belt will move
to one side and therefore the running belt needs to be readjusted.
N
Once the belt is readjusted, STOP the belt and operate the machine with the subject on the belt.
In most cases the readjustment of the running belt requires the same revolutions of the belt adjustment screw.
Therefore we recommend to note the required revolutions of the belt adjustment screw. This will help to avoid the
waiting time and can reduce the belt adjustment procedure to some seconds effort.
[12.N]
Bottle holder
The bottle holder is available for all h/p/cosmos running machines and
can be easily fixed with a clamp on the handrail. The position of the
bottle holder must not be at the rear end of the handrail due to this end
must be reachable for the subject.
bottle holder for handrail 60mm [cos11020]
[12.O]
Retrofitting UserTerminal
All lt-models can be retro-fitted with a UserTerminal. h/p/cosmos
oversized running machines, which are equipped with an external control
unit, can be complemented optionally with an additional UserTerminal at
the handrail.
In this case the external UserTerminal is connected via interface to the
running machine. To connect other devices, e.g. PC or ECG, an
additional interface (com 2 or com3) at the running machine is
necessary.