maintenance & safety inspections
file: n:\article\cos14310m5-v1_06hpc-en\20120215_cos14310m5-v1.06hpc-en_manual_h-p-cosmos_running_machine.doc
© 2012 h/p/cosmos sports & medical gmbh [email protected] created 15.02.2012 printed 15.02.2012 page: 129 of 213
[9.D2]
Lubricant test
The condition of the oil film between running surface and running belt can be established with the lubricant-test
[cos14379] which is available at h/p/cosmos. The lubricant-test consists of a pre-wheighed commercial paper
tissue. The tissue needs to be pulled through the gap between running belt and running surface in a defined lenght
in order to absorb a certain quantity of lubricant. Based on previous experience h/p/cosmos is able to evaluate the
quantity of the oil film by measuring the weight difference. However, it needs to be considered that this test can
only find out approximate values and trends and can never replace the direct survey by the authorized h/p/cosmos
technician in the prescribet time intervals.
For security reasons, the running machine needs to be switched off and the power plug needs
to be pulled out for full separation from the mains.
[9.D2]
Step Action
Reaction/display
[01]
There are 3 pre-weighed paper tissues
[cos14379] required for the test.
Every single tissue is located in a separate
plastic bag. The information (blue letters) in
the left picture need to be entered to the
sticker which is on the plastic bag.
[02]
10 cm
10 cm
Taschentuch
Tissue
Laufgurt / running belt
Trittfläche / footboard
L
R
1 meter
Trittfläche / footboard
M
The tissue-test needs to be performed at 3
places under the running belt:
-
Left edge
-
center
-
right edge
For each test a separate tissue needs to be
used!